G1362A Brechungsindexdetektor der Series 1100



The detector is designed for highest optical performance, GLP compliance and easy maintenance. It includes the following features:

  • advanced temperature controlled detector optics ready to use within two hours of installation
  • automatic zero and automatic purge combined with a recycle valve for automatic solvent recycling allow uninterrupted operation
  • durable tungsten lamp with a life expectancy of 40,000 hours
  • automatic light intensity control circuit to ensure the optimum performance of the optics
  • integrated diagnostics for efficient troubleshooting
  • built-in refractive index calibration
  • front access to valves and capillaries for easy maintenance

... mehr Details

Art.Nr.: G1362A_01

Gewicht: 12,00 KG

Lieferzeit 1 Woche | Sofort versandfähig, ausreichende Stückzahl
Auf den Merkzettel

Produktbeschreibung

Spezifikationen

Type
Specification Comments
Detection type Refractive Index
Refractive index range 1.00 - 1.75 RIU, calibrated
Measurement range
 +/- 600 x 10^5 RIU

Optical zeroing
via set screw
Optics temperature control
5 ° C above ambient to 55 ° C
Sample cell volume 8uL maximum pressure 5 bar (0.5Mpa) maximum flow rate 5mL/minute
Valves
Automatic purge and automatic solvent recycle

Volumes Inlet port to sample cell 62uL, inlet port to outlet port 590uL
Liquid contact materials
316 stainless steel, teflon and quartz glass

pH range 2.3 - 9.5
Performance specifications
Short term noise < +/- 2.5 x 10^9 RIU. Drift < 200 x 10^-9 RIU/hour see note below this table
Time programmable parameters polarity, peak width
Detector zero
automatic zero before analysis

Control and data evaluation
Parameter entry, signal display,on-line help and diagnostics with the Agilent 1200 Series Control Module. Optional PCMCIA card for method, sequence and logbook storage and transfer. Agilent ChemStation for LC PC based software for control and data evaluation.
Analog outputs Recorder/integrator: 100 mV or 1 V, output range selectable, one output
Communications Controller-area network (CAN), GPIB, RS-232C, LAN, APG Remote: ready, start, stop and shut-down signals
Safety and maintenance Extensive diagnostics, error detection and display (through control module and ChemStation), leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas
GLP features Early maintenance feedback (EMF) for continuous tracking of instrument usage with user-selectable limits and feedback messages. Electronic records of maintenance and errors. Automated operational qualification/performance verification (OQ/PV)
Housing All materials recyclable.
Environment 0 to 55 ° C constant temperature at <95% humidity (non-condensing)
Dimensions 180 mm x 345 mm x 435 mm (7 x 13.5 x 17 inches) (height x width x depth)
Weight 17 kg (38 lbs)